– 12 –
System Features
1. Listed UL1069 Patient to Staff Communication System for up to 80 stations.
2. Voice-actuated handset or press-to-talk communication.
Single and private channel (cannot be broken into).
3. Up to 8 trunk lines for subs. Max. 10 sub stations per trunk line.
3 twisted pair looped wiring, plus one separate wire for zone light indicator.
4. Compact sized 50-call master station.
30-call add-on selector simply plugs into master. (Expansion card required for CCU)
5. Single or dual-call Nurse Call bedside sub stations, able to have multiple stations in one room with
common bathroom pullcord and corridor lamp.
6. NHR-3TS Bathroom/Hallway urgent call sub station used with NHR-7A pullcord or switch.
7. Listed UL 1069 components: Corridor lamps, Bathroom pullcords,
urgent call switches, and Zone lights.
8. Three types of calls with distinct tone and light signalling.
1) Routine Call
2) Urgent Call- from bathroom common area bath or urgent call station
3) Cord-out trouble call from bedside.
Calls are answered in order of urgency and time of call.
9. Call tone & LED annunciation at Duty station.
10.Station number programming on each Sub station.
11.Sub stations connection verification at NHX-50M Master station.
12.Line-off trouble sounds intermittent buzzer with LED illumination.
13.Plug in wire terminations at Master and Sub stations. Makes replacement easy.
System capabilities
1.
A call from dual-call bedside sub (NH-2SA/A) lights the same call-in LED on the NHX-50M or NHX-30G.
2. NHX-50M master holds max. five call-in LED’s simultaneously.
3. The following features are not included in the NHX system;
• Second master in parallel (for routing calls). • Ceiling mount sub
• Desk or hand-held sub. • Call extension speaker
• Communication with 2 or more subs at a time. • Background music
• All call