Air Health AH-RL, AH-1 instruction sheet Maintenance, Replacement Procedure, Figure Figure

Models: AH-RL AH-1

1 16
Download 16 pages 6.78 Kb
Page 4
Image 4
MAINTENANCE

8.Wipe off the bulb glass with the supplied alcohol wipe. Then insert the bulb through the center hole of the mounting base. (See Figure 5) The power cord must be 90° to the view port hole on the mounting base.

9.With the ballast pressed against the mounting base, rotate the ballast clockwise 90°.

10.Plug the power supply cord into a standard electric wall outlet. Check bulb operation by looking through the view port hole. NOTE: The view port hole is made of glass and will not allow any UVC rays to pass through. A blue glow should be seen in the duct.

11.If blue glow is not seen, make sure base is properly installed onto duct. There is a proving switch located on the base that needs to be seated properly onto the duct. Make sure that

when base is mounted that the plated washer on the switch is raised approximately 3/16” and the switch is fully engaged. (See Figure 6)

12.With standard duct tape or like material, seal all joints and seams of the plenum near the

mounting location to reduce any direct viewing of light.

MAINTENANCE

WARNING: Unplug unit from outlet before commencing inspection and cleaning.

NOTE: Wear disposable, non-absorbent protective gloves prior to inspection procedure in case of accidental breakage of glass.

Figure 5

Figure 6

1.We recommend inspecting and cleaning bulb at least every 6 months depending upon your home environment and condition of conventional filters and duration of use and their replacement cycle.

2.To maintain the bulbs maximum output intensity , replace your Air+Health™ Air Purifying System bulb annually, depending upon your home environment using the replacement procedure below. The bulb should operate continuously for maximum bulb life and light effectiveness.

REPLACEMENT PROCEDURE

WARNING: Unplug unit from wall outlet before starting replacement procedure.

NOTE: Wear disposable, non-absorbent protective gloves prior to replacement procedure in case of accidental breakage of glass.

1.Rotate the Air+Health™ ballast assembly counter-clockwise until it stops. (See Figure 7)

2.Remove the ballast assembly and bulb from the mounting base. (See Figure 5)

3.Remove the bulb from the bulb socket. (See Figure 4)

4.Install replacement bulb into bulb socket. Do not handle the bulb with bare hands. Oils on the hands tend to reduce bulb intensity.

5.Wipe off the glass with the supplied alcohol wipe.

6.Position the ballast assembly power cord 90° to the view port hole on the mounting base. (See Figure 7)

7.Slide the bulb thought the hole in the mounting base. (See Figure 5)

8.Press the ballast against the mounting base and rotate the ballast assembly clockwise until it stops. (See Figure 8)

9.After assembly is firmly in place, plug the power supply cord into a standard electrical wall outlet. Check bulb operation by looking through the view port hole. A blue glow should be seen in the duct.

Figure 7

Figure 8

WARNING: Never expose eyes or skin to UVC light from any source. Looking directly at the UVC light may cause permanent Figure 5 eye damage or blindness. Never operate the Air+Health™ Air Purifying System out of the plenum. Avoid touching the REPLACEMENT PROCEDUREManual background glass portion of the bulb with your hands.

Page 4

Page 4
Image 4
Air Health AH-RL, AH-1 instruction sheet Maintenance, Replacement Procedure, Figure Figure