What You See

 

What It Means

What You Should Do

 

 

 

 

Control Test

Control solution is

Run another control test. If the

result is out of

 

past its expiration

result is still out of range, retest

range (too high

 

date or is past the

with a new 10-Test Disc and

or too low).

 

6-month open

control solution with an acceptable

 

use-life date.

expiration date and open use-life

 

 

 

10-Test Disc is past its

date. If still out of range, contact

 

the Bayer Diabetes Care Helpline.

 

 

expiration date.

 

 

 

 

Deteriorated test

 

 

 

strip due to heat or

 

 

 

exposure to moisture.

 

 

Control solution

 

 

 

is not at room

 

 

 

temperature.

 

 

• Not enough control

 

 

 

solution drawn into

 

 

 

test strip.

 

 

 

 

Meter begins

The inside of your

Wick the mosture away (see

countdown before

meter may be wet.

page 34). If you continue to

blood is applied.

 

 

experience difficulty, contact

 

 

 

the Bayer Diabetes Care Helpline.

 

 

 

Some display

This may affect the way

If there is a difference in your

segments do not

you see your results.

meter’s display and the one

appear to be

 

 

shown on page 5, contact the

showing when

 

 

Bayer Diabetes Care Helpline.

is held in or

 

 

 

the meter handle

 

 

 

is all the way out.

 

 

 

 

 

 

 

44

Bayer Diabetes Care Helpline: 1-800-268-7200

Page 48
Image 48
Bayer HealthCare 2 manual Control solution with an acceptable