Bayer HealthCare 2 manual What You See What It Means What You Should Do

Models: 2

1 53
Download 53 pages 28.29 Kb
Page 45
Image 45

What You See

What It Means

What You Should Do

 

 

 

 

 

 

 

 

Low Battery symbol

Replace the battery (CR2032)

 

 

 

flashes if battery power

within 20 readings or one week.

 

 

 

is low.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Battery life has ended.

Replace the battery (CR2032).

Temperature is outside

Allow the meter to reach room

the operating range of

temperature.

the meter (too hot or

 

too cold).

 

 

 

Meter does not

Remove the test disc and inspect

recognize the test disc.

the meter for foreign material.

 

Rotate the disc and reinstall it. If

 

error repeats, try inserting a new

 

disc. If error continues, contact the

 

Bayer Diabetes Care Helpline.

 

 

Test strip was removed

Take care to not press the release

during the test, or the

button or remove the test strip

release button was

during the test. Release the used

pressed during the test.

test strip and repeat the test

 

(see page 10). If the error contin-

 

ues, contact the Bayer Diabetes

 

Care Helpline.

The meter was open

Close the meter and push the

when the meter handle

handle in.

was pulled.

 

(with continuous beep)

www.ascensia.ca41

Page 45
Image 45
Bayer HealthCare 2 manual What You See What It Means What You Should Do