Braun BP3560, BP3550 manual What to do if, Problem Reason Solution

Models: BP3550 BP3560

1 74
Download 74 pages 12.8 Kb
Page 12
Image 12

What to do if ...

Problem

Reason

Solution

 

 

 

 

 

 

 

Batteries are flat.

Insert new batteries.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Memory readings can be

 

 

 

 

 

 

 

 

called up.

 

 

 

 

 

 

 

 

Measurements are not

 

 

 

 

 

 

 

 

possible.

 

(flashing)

 

e.g.

Error codes 01-29:

Refit device and repeat

................

 

Device may not be

measurement as shown and

 

correctly positioned on

described under «Measuring».

 

 

 

 

 

 

 

the wrist.

 

 

 

 

 

 

 

 

 

 

e.g.

Error codes 30 and

Repeat measurement. If problem

................

 

higher:

persists, have device checked at a

 

System error

Braun Service Centre.

 

 

 

 

 

 

 

 

 

Measurement

Right measuring position

Repeat measurement as shown

or inflation

was not reached.

and described under «Measuring»

does not start

 

or deactivate the positioning

instantly.

 

system (A).

 

 

 

 

 

 

 

 

 

Readings from

Blood pressure is a

none

repeated mea-

fluctuating value.

 

surements differ

For healthy adults,

 

considerably.

deviations of about 10 to

 

 

 

 

 

 

 

 

20 mmHg are possible.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cuff is not properly fitted.

Ensure that the appliance is fitted

 

 

 

 

 

 

 

 

as follows:

 

 

 

 

 

 

 

 

– on top of the slim side of the

 

 

 

 

 

 

 

 

forearm

 

 

 

 

 

 

 

 

– with the display on top

 

 

 

 

 

 

 

 

– with the battery compartment

 

 

 

 

 

 

 

 

facing the user’s body

 

 

 

 

 

 

 

 

– approx. 2 cm away from the

 

 

 

 

 

 

 

 

thumb root

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Readings were not con-

For each measurement, proceed

 

 

 

 

 

 

 

sistently taken at heart

exactly as shown and described

 

 

 

 

 

 

 

level.

under «Measuring».

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Talking, coughing,

While measuring, relax, keep still,

 

 

 

 

 

 

 

laughing, moving etc.

do not move, or speak.

 

 

 

 

 

 

 

when measuring will

 

 

 

 

 

 

 

 

influence the reading.

 

 

 

 

 

 

 

 

 

 

Prior to a

Positioning system (A) is

Refer to the chapter «Select mode»

measurement,

deactivated.

to activate the positioning system

the positioning

 

(A).

system (A) on the

 

 

display does not

 

 

work.

 

 

 

 

 

 

 

 

 

 

 

12

Page 12
Image 12
Braun BP3560, BP3550 manual What to do if, Problem Reason Solution