Product specifications
Displayed temperature range: | 93.2 °F – 108 °F | (34 °C – 42.2 °C) |
Operating ambient |
|
|
temperature range: | 50 °F – 104 °F | (10 °C – 40 °C) |
Display resolution: | 0.1 °F or °C |
|
Accuracy for patient temperature range | Maximum Laboratory Error | |
96.8 °F – 102.2 °F (36 °C – 39 °C): | ± 0.4 °F | (± 0.2 °C) |
outside this range: | ± 0.5 °F | (± 0.3 °C) |
Long term storage ranges |
|
|
Temperature: | ||
Humidity: | 95 % | |
Battery life: | 2 years / 1000 measurements |
This infrared thermometer meets requirements established in ASTM Standard E
ASTM laboratory accuracy requirements for the thermometer only (not including a lens filter) in the display range of 96.8 °F to 102.2 °F (36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas for
This appliance conforms to the following standards: DIN EN
DIN EN
Part 5: Performance of infrared ear thermometers (with maximum device)
TYPE BF | Attention, consult |
EQUIPMENT | ACCOMPANYING |
| DOCUMENTS |
Internally Powered Equipment
Continuous Operation
Not Protected against Ingress of Water
U.S. Patent No. 5,088,834 Other Patents Pending
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact the Braun ThermoScan customer service at
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.