KEY TO DISPLAY ICONS

Icon displayed

Condition/Cause

Corrective action

 

 

 

 

 

 

Insuffi cient pressure.

Repeat measurement

 

 

 

again.

 

 

 

 

 

 

Appears when the blood-

Reapply cuff properly and

 

 

pressure could not be

measure again.

 

 

measured accurately.

 

 

 

 

 

 

 

The pressure is 281

If the pressure does not

 

 

mmHg or higher.

automatically decline

 

 

 

 

 

during measurement,

 

 

 

immediately press the

 

 

 

 

 

 

“START/STOP” switch

 

 

 

and turn the unit off. To

 

 

 

take the measurement

 

 

 

again, refer to P.17

 

 

 

and correctly re-apply

 

 

 

pressure.

 

 

 

 

 

 

Appears when the

Replace the batteries.

 

 

batteries are running low.

 

 

 

 

 

 

 

Detected problem in the

Ask purchased store.

 

 

device.

 

 

 

 

 