KEY TO DISPLAY ICONS

English

Icon displayed

 

Condition/Cause

 

 

 

 

 

Displayed during

Lit

Measurement is under way.

 

 

 

 

measurement.

Flashing

Pulse detected.

 

 

 

 

 

 

 

Displayed during

“PUL Err” is displayed when the pulse

 

measurement.

value is below 40 or above 181.

 

 

 

 

If any of the following marks is displayed, take the specified action, then repeat the operation.

Icon displayed

 

Condition/Cause

Corrective action

 

 

 

 

 

Insufficient pressure.

Apply cuff properly and

 

Appears when the

measure again.

 

 

blood-pressure could

 

 

 

not be measured

 

 

 

accurately.

 

 

 

 

 

The pressure is 281

If the pressure does not

 

mmHg or higher.

automatically decline

 

 

 

during measurement,

 

 

 

immediately press the

 

 

 

“START/STOP” switch

 

 

 

and turn the unit off. To

 

 

 

take the measurement

 

 

 

again, refer to P.16

 

 

 

and correctly re-apply

 

 

 

pressure.

 

Appears when the

Replace the batteries.

 

batteries are running low.

 

 

 

 

 

Detected problem in the

Contact the vendor.

 

device.

 

 

 

 

 