SPECIFICATIONS

Measuring Range:

For Ear temperature mode 34.0~42.2°C

 

(93.2~108°F)

 

 

 

For Material surface temperature mode -22~80°C

Accuracy:

(-7.6~176°F)

 

 

For Ear temperature mode

 

±0.2°C (0.4°F) for temperatures 35.5~42.0°C

 

(95.9~107.6°F). ±0.3°C (0.5°F) for other ranges.

 

For Material surface temperature mode

 

±0.3°C (0.5°F) for temperatures 22~42.2°C

 

(-7.6~108°F) for other ranges. ±4% or ±2°C (4°F),

Display:

whichever is greater.

3 digits with decimal point

 

Battery mark

 

 

 

 

 

 

 

 

 

Ear temperature mode

 

Material surface temperature mode

 

Stopwatch

 

 

 

 

 

Memory

 

 

 

 

Battery:

°C or °F

 

 

 

 

One lithium cell battery (CR2032 X 1 pc)

Battery life:

3000 consecutive measurements or 1 year with 1-2

Rated Voltage:

measurements per day including stand-by mode.

DC3V

 

 

(

 

 

: Direct Current)

 

 

 

Operating temperature range:

10~40°C (50~104°F)

Storage temperature range:

-20~50°C (-4~122°F)

 

Transportation temperature range must be less than

Unit size:

70°C.

 

 

 

 

115 (L) X 37 (W) X 29 (H) mm

Weight:

Approx. 60g (including battery)

Electric shock protection: Internal power unit

Safety classification:

Type BF equipment

Memory:

9 memory slot recall function

Attention, consult accompanying documents.

*Specifi cations may be changed without notice in the event of improvements being made.

Reference standard: Complies with ASTM E1965-98, EN12470-5: 2003

Clinical thermometers-Part 5: Performance of infra-red ear thermometers (with maximum device), IEC60601-1-2(EMC), EN60601-1(Safety) standards.

Manufacturer:

Name: Radiant Innovation Inc.

Address: 1F, No.3, Industrial East 9th Road, Science-Based Industrial Park, HsinChu, Taiwan 300.

European Representative:

Name: Medical Technology Promedt Consulting GmbH

Address: Altenhofstrasse 80, D-66386 St. Ingbert, Germany

English