Drive Medical Design 14027 Always read the operating instructions before use, Operation, Step

Models: 14027

1 20
Download 20 pages 23.99 Kb
Page 11
Image 11
!Always read the operating instructions before use.

OPERATION

!Always read the operating instructions before use.

Please follow instructions below for detailed operating procedure of each type.

Step 1

Turn on the power. The indicator light on the power switch will come on.

The pump starts to pump air into the mattress.

The Low Pressure indicator (yellow or red) will come on as inflation undergoes.

The audible alarm is activated, accompanying with the Low Pressure indicator, to alert that pressure is low. Press the Alarm Reset button to mute the audible alarm, and its indicator will flash.

NOTE! For first inflation, we recommend that you press both the Alarm Reset and Static buttons to enter a quick and silent operation.

Step 2

When the pressure reaches the preset level by factory default within approximately 30~45 minutes, the Normal Pressure indicator will come on, and the Low Pressure indicator (and the audible/visible alarm) will turn off.

NOTE! If the Low Pressure indicator (and the audible/visible alarm) won’t go off, please refer to page 15 for troubleshooting.

Step 3

Turn the pressure-adjust knob to set a comfortable pressure level from soft to firm. (Pressure range: 30~60 mmHg for Q2-02).

NOTE! For suitable pressure, please refer to page 9 for the hand check procedure.

Step 4

The pump is preset in alternating mode and its cycle time is set at 10 min/60Hz or 12 min/50Hz.

Press the Static button to set it in static mode, and the Static indicator will come on. The static mode will be started within approximately 6 minutes. Press the Static button again to switch back to alternating mode.

NOTE! In static mode, the mattress provides a firm surface that makes it easier for the patient to transfer or reposition. The static mode prevents the patient from bottoming out when in a sitting position.

Step 5

During normal operation, the pump will monitor pressure changes. If the pressure is below a user- defined pressure level, the pump will automatically start to inflate the mattress. Accompanying with the Low Pressure indicator, the audible alarm will beep and its indicator will come on to bring attention to a low pressure condition. The pump will stop when the user-defined pressure level is reached.

Press the Alarm Reset button to mute the alarm; its indicator will flash.

Press the Alarm Reset button again to have the audible alarm and its indicator on.

11

Page 11
Image 11
Drive Medical Design 14027 manual Always read the operating instructions before use, Operation, Step