5cleaning and maintenance
Nebulizer Cleaning
All parts of the nebulizer, except tubing, should be cleaned according to the following in- structions. Your physician and/or Drive provider may specify a certain cleaning procedure. If so, follow their recommendations.
To prevent possible risk of infection from contaminated medication,
cleaning of the nebulizer is recommended after each aerosol WARNING! treatment. Disinfecting is recommended once a day.
Clean After Every Use:
1.With power switch in the “ Off” position, unplug power cord from wall outlet.
2.Disconnect tubing from the
3.Disassemble mouthpiece or mask from cap. Open nebulizer by turning cap counterclockwise and remove baffle.
4.Wash all items, except tubing, in a hot water/dishwashing detergent solution. Rinse under hot tap water for 30 seconds to remove detergent residue. Allow to air dry.
5.For reusable nebulizer only, it can be cleaned daily in dishwasher using top shelf or by following the above cleaning procedure.
The Drive Reusable Nebulizer is dishwasher safe, but do not place
any Drive Disposable Nebulizer parts in an automatic dishwasher; CAUTION! doing so could cause damage.
Disinfect Daily
1.Using a clean container or bowl, soak items in three parts hot water to one part white vinegar for 30 minutes OR use a medical
2.With clean hands, remove items from disinfectant solution, rinse under hot tap water, and air dry on a clean paper towel. Store in a
NOTE | Do not towel dry nebulizer parts; this could cause contamination. |
To prevent possible risk of infection from contaminated cleaning
solutions, always prepare fresh solution for each cleaning cycle WARNING! and discard solution after each use.
................................................................................................................ 11
DRIVE PACIFICA ELITE COMPRESSOR/NEBULIZER MODEL 18070
................................................................................................................