3.Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does not have to be washed because only filtered air passes through it.

NOTE

The Drive Disposable Nebulizer will last at least 15 days and possibly

 

longer, depending on usage. Proper cleaning will also help extend the life

 

of the nebulizer. Because it is disposable, we recommend that an extra

 

nebulizer be kept on hand at all times. The Drive Reusable Nebulizer is

 

dishwasher safe and may be cleaned and reused for up to one year.

Compressor Cleaning

1.With power switch in the “Off ” position, unplug power cord from wall outlet.

2.Wipe outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.

Do not submerge in water; doing so will result in compressor damage.

WARNING!

 

 

Filter Change

1.Filter should be changed every 6 months or sooner if filter discolors.

2.Remove filter by grasping it firmly and pulling out of the unit. Discard old filter.

3.Replace with a new Drive filter. Additional filters should be purchased from your Drive provider.

4.Push new filter into position.

Filter cover

Do not reuse or wash filter or substitute any other material such as

 

WARNING!

cotton for a Drive air-inlet filter. It will result in compressor damage.

 

 

 

Maintenance

All maintenance must be performed by a qualified Drive provider or authorized service center.

Electric shock hazard. Do not remove compressor cabinet. All

DANGER!

 

disassembly and maintenance must be done by a qualified Drive

 

 

provider.

 

12 .............................................................................................................

DRIVE PACIFICA ELITE COMPRESSOR/NEBULIZER MODEL 18070

.............................................................................................................

Page 12
Image 12
Drive Medical Design 18070 manual Compressor Cleaning, Filter Change, Maintenance, Filter cover