Drive Medical Design 18605 user manual cleaning &maintenance

Models: 18605

1 16
Download 16 pages 13.35 Kb
Page 10
Image 10
cleaning &maintenance

cleaning &maintenance

9

 

Collection Bottle:

1.Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the DC input receptacle on the unit.

2.With the collection bottle still in the holder, remove the lid. The bottle can now be taken out of the holder to be emptied.

NOTE

Collection bottle should be emptied and cleaned after each use. The 800ml disposable collection bottle and lid are meant for single-patient use

3.Collection bottles and lids should be thoroughly cleaned after each use by one of the following methods:

All Bottles:

. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry. Autoclave for 3 to 5 minute cycle at 270oF (132oC).

In lieu of autoclave sterilization, the above parts can be soaked in an activated dialdehyde solution.

Suction Unit:

1.With the power switch in the “OFF” position, disconnect the Drive Suction Unit from all exter- nal power sources.

2.Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.

CAUTION

Do not submerge in water as this will result in damage to the vacuum pump.

Tubing:

1.Disconnect the tubing from the unit.

2.Tubing should be rinsed thoroughly after every use by running hot tap water through it fol- lowed by a solution of one part vinegar to three

parts hot water.

3.Rinse with hot tap water and air dry.

4.Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.

Carrying Case:

1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.

Page 10
Image 10
Drive Medical Design 18605 user manual cleaning &maintenance