3. INSTALLATION

3. INSTALLATION

Unpack the box and check the package contents for completeness.

Package Content List

nMattress unit x 1 (may not be included if only pump unit is purchased)

nPump Unit x 1

nUser Manual x 1

Inspect equipment for damage which may have occurred during shipment. If there is damage, please contact your dealer immediately.

3.1 PUMP & MATTRESS INSTALLATION

1.Place the mattress or pad on top of a bed frame. Orient the mattress so that the foot-markers on the mattress are situated towards the end of the bed (where pump will be located).

2.Place the pump hangers over the bed railing at the foot of the bed. Support the pump from the bottom and turn the hangers outward (or inward) to secure the pump against the railing.

If no bed railing is available, place the pump on a flat, sturdy surface or on the floor, beneath the bed.

3.Connect the Quick Connector on the air hose (from air mattress) to the pump unit Quick Connector Slot. When you hear a "click", the connection will be secured.

NOTE: Check and ensure the air hoses are not kinked or tucked under the mattress.

8. TECHNICAL DESCRIPTION

Specifications:

 

Item

 

 

Specification

 

Power Supply (Note: See

 

 

AC 100-120V 60 Hz, 0.16 A 14W (For 120V System)

 

rating label on the product)

 

 

 

 

 

 

 

 

Fuse Rating

 

T1A, 250V

 

 

Cycle time

 

Fixed, 8 min @ 60Hz,

 

 

Dimension (L x W x H)

 

29 x 18.5 x 12.6 (cm) or 11.4" x 7.3" x 5.0"

 

 

Weight

 

2.2 Kg or 5 lb

 

 

 

 

 

 

Operation:10° C to 40° C (50° F to 104° F)

 

 

 

Temperature

 

 

Storage: -15° C to 50° C (5° F to 122° F)

 

Environment

 

 

 

Shipping: -15° C to 70° C (5° F to 158° F)

 

 

 

 

Operation: 10% to 90% non-condensing

 

 

 

 

 

 

 

 

 

Humidity

 

 

Storage: 10% to 90% non-condensing

 

 

 

 

 

 

Shipping:10 % to 90% non-condensing

 

 

 

 

 

 

Class II with functional earth(NOTE:5*), Type BF, IPX0

 

 

Classification

 

 

Applied Part: Air Mattress

 

 

 

 

Not suitable for use in the presence of a flammable

 

 

 

 

 

 

 

 

 

 

 

 

anesthetic mixture (No AP or APG protection)

 

 

Mattress

 

 

Specification

 

 

Model

 

 

8” Replacement

 

 

Dimension (L x W x H)

 

 

200 x 90 x 20.3 (cm)

 

 

 

 

or 78.7” x 35.4” x 8”

 

 

 

 

 

 

 

 

Weight

 

 

7.5 kg 16.5 lb

 

 

Pressure Range

 

 

Adjustable Weight Settings

 

Package Content List NOTE: 1. Consult the distributor or EU representative for further technical documents.

2.These specifications are also applicable for other regions operating with the same power supply.

3.The manufacturer reserves the right to modify the specifications without notice.

4.The earthing terminal is served as functional earth, not protective earth.

5

10

Page 6
Image 6
Drive Medical Design AS 5000 Technical Description, Pump & Mattress Installation, Package Content List, Specification