4. OPERATION
NOTE: Always read the operating instruction before use.
4.1 GENERAL OPERATION
1. Switch on the main power switch found from the
side of the pump.
2. The pump unit is to delivering air into mattress, wait
least 20 minutes for mattress to be inflated.
3. The low-pressure indicator (yellow LED) will light up
when the mattress is not fully inflated.
4. When the appropriate pressure is reached, the
low-pressure indicator (yellow LED) will go off and
pump will enter the alternating mode.
. According to the weight and height of the patient,
adjust the pressure setting to the most comfortable
level without bottoming out, then the pressure in
mattress will slowly increase to a certain value after
the air mattress is ready to use.
NOTE: Every time when mattress is first setup
for use, it is recommended that the
pressure knob to be set to "Max" for
the quickest inflation. User can then
adjust air mattress to the desired
softness afterward.
4.2 CPR
When there is an emergency to perform CPR on the
patient, quickly pull the CPR valves to release air from
mattress. The CPR valves are located at the head-end,
right-hand-side of the mattress. The quick connector
found from the pump unit can be disconnected for
even faster deflation.
9
2. Cycle Time
There are four selectable cycle times found from the touch panel area. By pressing (icon) sign,
user can select four different cycle times for best comfort.
3. Therapy.
A. Autofirm
This mode will be forced to execute every time the power switch is turned on. This insures the
pump is able to reach its maximum operating pressure. Once the max pressure level is
reached, the pump will automatically switch into alternating mode. User can also use this
function as full mattress inflation during ingress/egress the patient for better support.
B. Alternate
There are four selectable cycle time available. User can select for best comfort.
C. Static.
Press STATIC button to suspend alternating function, if needed. The pressure inside of air
tubes will be adjusted to the same softness. Press the STATIC button again; it will switch back
to alternating mode. Under the static mode, cell pressure level will be lowered compare to the
same pressure level from alternating mode. This special feature is to protect high-risk
pressure sore patient from receiving additional pressure, which could lead to pressure sore
risks. There is a 20 mins auto-recovery mode back to alternating cycle to ensure patient
always getting best therapy possible.
D. Seat Inflation
The seat inflation features additional supports to the patient during upright position without
bottoming out. User can select this additional feature under either static or alternate mode.
4. Panel Lock
Should the panel remain untouched for 3 minutes, the panel lock feature will lock the panel with
green LED light on to prevent accident from changing the setting during normal operation. To
unlock, simply by pressing the Panel lock button for 3 second to unlock.
5. Alarm Mute
Press alarm mute button to temporary suspend the low-pressure alarm (LED light and buzzer).
Should the situation not resolved within 5 mins, the alarm shall resume to notify the patient.
A. PFA (Power Failure Alarm)
During power failure situation, the Power Failure LED light will lit on with buzzer. By pressing
the mute button to disable both buzzer and LED.
B. Low Pressure Indicator
When low pressure LED lights up, the pressure inside of air mattress is below normal. Please
refer to troubleshooting.
6