MODEL 140,141, and 142 sensors

TROUBLESHOOTING

TROUBLESHOOTING

PROBLEM

PROBABLE CAUSE

SOLUTION

 

 

 

Off-scale reading

Wiring is wrong.

Verify wiring.

 

 

 

 

 

Temperature element is open or shorted.

Check temperature element for open or

 

short circuits. See Figure 14.

 

 

 

 

 

 

Sensor is not in the process stream.

Be sure the sensor is completely submerged

 

in the process stream.

 

 

 

Sensor has failed.

Perform isolation checks. See Figure 15.

Noisy reading

Sensor is improperly installed in the

Be sure sensor is completely submerged

 

 

process stream.

in process stream.

 

 

 

Reading seems wrong

 

Be sure the sensor is properly oriented in the

 

Bubbles trapped in the sensor.

(lower or higher than

pipe or flow cell. See Figure 1. Apply back

 

 

expected)

 

pressure to the flow cell.

 

 

 

 

Check that the temperature correction is

 

Wrong temperature correction algorithm.

appropriate for the sample. See analyzer

 

 

manual for more information.

 

 

 

 

 

Verify that the correct cell constant has

 

 

been entered in the analyzer and that the

 

Wrong cell constant.

cell constant is appropriate for the

 

 

conductivity of the sample. See analyzer

 

 

manual.

Sluggish response

Electrodes are fouled.

Clean electrodes.

 

Sensor is installed in a dead area in the

Move the sensor to a location more

 

piping.

representative of the process liquid.

 

 

 

 

 

 

 

 

 

 

 

 

orange

 

 

 

 

 

 

 

 

 

RTD

 

 

 

 

 

 

 

 

red

R

 

 

 

 

 

 

 

 

 

 

 

 

 

 

red

 

 

 

 

 

 

 

 

 

 

 

 

 

 

terminal strip in sensor

 

 

gray

 

 

 

 

 

 

 

junction box

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

temperature

resistance in :

 

θC

Pt 100

 

10K NTC

 

100K NTC

0

 

 

 

 

100.0

 

29.49K

 

351K

10

 

 

 

 

103.9

 

18.79k

 

208K

20

 

 

 

 

107.8

 

12.26K

 

127K

30

 

 

 

 

111.7

 

8194

 

 

79.4K

40

 

 

 

 

115.5

 

5592

 

 

51.0K

50

 

 

 

 

119.4

 

3893

 

 

33.6K

FIGURE 14. Checking Temperature Element

Disconnect cable leads and measure resistances shown. The measured resistance should be close to the value in the table.

FIGURE 15. Checking Continuity and Leakage

Disconnect cable leads and measure resistance and con- tinuity as shown. Sensor must be dry when checking resistance between electrode leads.

11

Page 11
Image 11
Emerson 140, 141, 142 instruction sheet Troubleshooting, Problem Probable Cause Solution