Emerson 228 specifications Maintenance, Troubleshooting, Problem Probable Cause Solution

Models: 228

1 16
Download 16 pages 56.15 Kb
Page 14
Image 14

MODEL 228

MAINTENANCE

MAINTENANCE

WARNING

Retraction chamber contents may be under pressure.

CAUTION

Be sure sensor has been cleaned of process liquid before handling.

Generally, the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning frequency is best determined by experience.

TROUBLESHOOTING

PROBLEM

PROBABLE CAUSE

SOLUTION

 

 

 

Off-scale reading

Wiring is wrong.

Verify and correct wiring.

 

 

 

 

RTD is open or shorted.

Check the RTD for open or short circuits.

 

 

See Figure 19.

 

 

 

 

Sensor is not in process stream.

Confirm that the sensor is fully submerged in

 

 

the process stream. See installation section.

 

 

 

 

Sensor is damaged.

Perform isolation checks. See Figure 19.

 

 

 

Noisy reading

Sensor is improperly installed in the

Confirm that the sensor is fully submerged

 

process stream.

in the process stream. See installation section.

 

 

 

 

Sensor cable is run near high voltage

Move cable away from high voltage conductors.

 

conductors.

 

 

 

 

 

Sensor cable is moving.

Keep sensor cable stationary.

 

 

 

Reading seems wrong (lower

Bubbles are trapped in the sensor,

Install the sensor in a vertical pipe run with the

or higher than expected)

particularly in the toroid opening.

flow against the toroid opening. Increase flow

 

 

if possible.

 

Sensor is not completely submerged in the

Confirm that the sensor is fully submerged

 

process stream.

in the process stream. See installation section.

 

 

 

 

Cell constant is wrong. Wall effects are

Calibrate the sensor in place in the process

 

present.

piping. See calibration section.

 

 

 

 

Wrong temperature correction algorithm

Check that the temperature correction is

 

is being used.

appropriate for the sample. See analyzer

 

 

manual for more information.

 

 

 

 

Temperature reading is inaccurate.

Disconnect the RTD leads (Figure 19) and

 

 

measure the resistance between the in and

 

 

common leads. Resistance should be close

 

 

to the value in Table 1.

 

 

 

 

Slow temperature response to sudden

Use an RTD in a metal thermowell for

 

changes in temperature.

temperature compensation.

 

 

 

Sluggish response

Sensor is installed in a dead area in the

Move sensor to a location more representative

 

piping.

of the process liquid.

 

 

 

 

Slow temperature response to sudden

Use an RTD in a metal thermowell for

 

changes in temperature.

temperature compensation.

 

 

 

14

Page 14
Image 14