MODEL 399VP pH/ORP SENSOR | SECTION 4.0 |
|
4.3ORP CALIBRATION
Most industrial applications have a number of ORP reactions occurring in sequence or simultaneously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP poten- tial is absolute because it is the result of the
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the direc- tions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in con- tact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
1.Make a temporary electrical connection between the sensor and the instrument.
2.Obtain an ORP standard solution (PN
3.Immerse the sensor in the standard solution. Allow
4.Adjust the standardize control of the instrument to the solution value shown on the label of the PN
5.Remove the sensor from the buffer, rinse, and install in the process.
TABLE
|
| pH 4 Solution | pH 7 Solution |
| ||||
|
|
|
|
|
|
|
|
|
| Temp °C | 20 | 25 | 30 | 20 | 25 | 30 |
|
|
|
|
|
|
|
|
|
|
| Millivolt Potential | 268 | 264 | 260 | 94 | 87 | 80 |
|
|
|
|
|
|
|
|
|
|
20