5.1 Analyzer Module Setup

5.1.1 Calibration Parameters – Calibration

Menu "Calibration Procedure Status":

Press the 8 -key or the -key in the line "Calibration procedure status..." to change to the corresponding submenu:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

 

 

 

-- Calibration Procedure Status --

 

 

 

 

 

 

 

 

 

 

 

Procedure status:

 

 

 

 

 

 

 

Ready

 

 

 

 

 

 

 

 

Remaining procedure time:

 

 

 

 

 

0 s

 

 

 

 

 

 

 

 

Current/expected gas flow:

 

 

 

Samplegas

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Concentration in span gas units:

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

----------------------Last zero calibration:

 

Results

----------------------

 

 

 

Success

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Last span calibration:

 

 

 

 

 

 

 

Success

 

 

 

 

 

 

 

 

Last zero calibration was:

Fri 05-29-1998 13:32:06

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Last span calibration was:

Fri 05-29-1998 13:37:23

 

 

 

 

 

 

 

 

Successful zero+span calibrated ranges:

1+2+3+4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Cancel !

 

 

 

 

Back...

 

More...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Calibration Procedure Status" you will find the results of the last calibrations. This menu will appear automatically after starting the zero or span gas calibration. Then it will display the status of the running calibration procedure (see sections 4.5/4.6 p. 4-39 to 4-46). If a calibration is still running, you can cancel it at any time with the F2 key.

With the F5 key (More...) you will change to a submenu, where you will find the calibration deviations of the current channel. The deviations of the other channels of an MLT analyzer or analyzer module you will find there by changing the channel with the F3 key.

TAG

 

 

 

37.50 ppm

 

-- Calibration Deviations --

 

 

Deviation from zero:

 

 

0.34

ppm

Sum of zero deviations:

 

 

4.41

ppm

Deviation from span:

 

 

0.82

ppm

Sum of span deviations:

 

 

7.57

ppm

Measure

 

Channel

Back...

 

 

F1

F2

F3

F4

F5

 

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 17