Notes:

The signal has to be stable before starting any calibration !

If you want to start the zero and the span calibration separately, or

if you want to calibrate separately each channel of an MLT analyzer/analyzer module, or if you want to calibrate the measurement ranges separately with span gas,

you have to change to the menu "Analyzer Module Calibration" (see p. 5-15)resp. to the menu "Calibration Procedure Setup" (p.5-9).

You can cancel any calibration procedures with the function "Cancel all running procedures !"

Via the F2 key you will reach the menu "Analyzer Channel Status". In that menu and its corresponding submenus you will find the status report of the current channel for

Failures

Maintenance requests

Function controls

Measurements/Alarms

Events

Hours of operation

and the operational settings of the current channel like

range settings and

response time (t90-time).

The menu "Analyzer Channel Status" and its submenus are displayed detailed in section 4.1.

Menu line "Info...":

Via this menu line you can change to the submenu "State of Calibration Procedures", where you will find the actual status of each channel of an MLT analyzer or analyzer module.

TAG

95.00 ppm

-- State of Calibration Procedures --

Channel -1:

 

 

 

Ready

Channel -2:

 

 

 

Ready

Channel -3:

 

 

 

Ready

Channel -4:

 

 

 

Ready

Channel -5:

 

 

 

Ready

Measure

 

 

Back...

 

F1

F2

F3

F4

F5

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 19