5.1 Analyzer Module Setup

5.1.1 Calibration Parameters – Advanced calibration methods

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

Calibration parameters

Advanced calibration methods

TAG

95.00 ppm

-- Advanced Calibration Methods --

Start zero calibration procedure for all channels !

Start span calibration procedure for all channels !

Start zero+span calibration procedure !

Start zero+span calibration procedure for all channels !

Start basic calibration procedure !

Cancel all running procedures !

Info...

Measure

Status...

 

Back...

 

F1

F2

F3

F4

F5

In the menu "Advanced Calibration Methods" you can

¨start the zero calibration of all measurement ranges for all channels of an MLT analyzer or analyzer module simultaneously or

¨start the span calibration of all measurement ranges for all channels of an MLT analyzer or analyzer module simultaneously or

¨start the zero and span calibration together for the current channel of an MLT analyzer or analyzer module or

¨start the zero and span calibration together for all channels of an MLT analyzer or analyzer module or

¨set all calibration deviations to zero ("basic calibration procedure").

Starting the functions:

¨Change to the line you want with the -key or the -key.

¨Press the 8 -key or the -key or

cancel and go back to the last menu page with the F4 key (Back...) or the -key.

¨If asked: Confirm with the F2 key (Yes) or

cancel and go back to the last menu page with the F4 key (Back...) or the -key.

5 - 18

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98