5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Advanced calibration methods
Main Menu — Analyzer and
↓
Analyzer module setup
↓
Calibration parameters
↓
Advanced calibration methods
↓
TAG | 95.00 ppm |
Start zero calibration procedure for all channels !
Start span calibration procedure for all channels !
Start zero+span calibration procedure !
Start zero+span calibration procedure for all channels !
Start basic calibration procedure !
Cancel all running procedures !
Info...
Measure | Status... |
| Back... |
|
F1 | F2 | F3 | F4 | F5 |
In the menu "Advanced Calibration Methods" you can
¨start the zero calibration of all measurement ranges for all channels of an MLT analyzer or analyzer module simultaneously or
¨start the span calibration of all measurement ranges for all channels of an MLT analyzer or analyzer module simultaneously or
¨start the zero and span calibration together for the current channel of an MLT analyzer or analyzer module or
¨start the zero and span calibration together for all channels of an MLT analyzer or analyzer module or
¨set all calibration deviations to zero ("basic calibration procedure").
Starting the functions:
¨Change to the line you want with the −
¨Press the 8
cancel and go back to the last menu page with the F4 key (Back...) or the ←
¨If asked: Confirm with the F2 key (Yes) or
cancel and go back to the last menu page with the F4 key (Back...) or the ←