NGA-2000 System Calibration

2 Valves for system calibration

Before using the system calibration we have to put together the required valve pool. There are in principle three types of I/O modules supporting this:

DIO

- 24 digital outputs / 8 digital inputs (max. 4 DIO modules per platform or 2 per

 

 

MLT/TFID analyzer)

SIO

- 3 digital outputs (max. 1 SIO module per CM)

[CVU - - 4 digital outputs (max. 4 modules per CM)]

Software supports up to 32 system valves. SIO and DIO are available, the control valve unit (CVU) is in progress.

2.1 Assigning an output port to a system valve

Assigning of an output can be accomplished by using the menus for selected output module (DIO, SIO or CVU).

There we have to select the NGA Control Module as the Source Module. Control module then provides the signals for system valve V1...V32.

For example the DIO:

Analyzer and I/O-module expert configuration...

DIO module(s)...

- DIO-MODULE OUTPUTS-

 

Inputs...

 

 

 

 

DIOOUTNUMC

Choose module...

 

 

Choose signal...

 

 

Module status:

NORMAL

DIOOUTINVC

DIOMODSTAC

Slot ID:

1

DIOSLOTIDC

Signal level:

OFF

DIOOUTSIGC

DIOOUTSTATC

Signal comes from:

NGA Control Module

DIOOUTSRCC

 

 

 

There we have to

select "Outputnumber"

then choose „NGA Control Module“ as module

Supplement - 4

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98