NGA-2000 System Calibration

Example for an NGA analyzer system:

 

V1

 

 

AM1 (MLT ch.1)

 

AM2 (MLT ch.2)

 

AM3 (MLT ch.3)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

V2

 

 

AM4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

V3

 

 

AM5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

V 4

 

V 5

 

 

V 6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration gas mixtures

 

 

 

Appropriate assignment:

 

 

 

 

 

 

 

 

 

AM1/Ch.1

 

 

AM2/Ch.2

AM3/Ch.3

AM4

AM5

SAMPLE-GAS

 

V1

 

 

 

V1

V1

V2

V3

ZERO-GAS

 

V4

 

 

 

V4

V5

V5

V6

SPANGAS-1

 

V5

 

 

 

V5

V4

V6

V5

SPANGAS-2

 

V5

 

 

 

V5

V4

V6

V5

SPANGAS-3

 

V6

 

 

 

V5

V4

V4

V4

SPANGAS-4

 

V6

 

 

 

V5

V4

V4

V4

Attention: Take in mind that any MLT channel is regarded as AM in this case!

Take care that for TFID separate valves are chosen which are not related to other AM’s or MLT channels (internal TFID valves only)!

This assigning procedure is to perform for all analyzer modules which should be included in system calibration.

To display the assigned valves and purge times for each AM we can push the softkey „View“.

If we want to exclude an AM from SYSCAL we can do this by entering an invalid valve. We can watch in the display whether an AM/channel is enabled for system calibration.

Attention: Remember to assign a system valve to an output-port!