Example for an NGA analyzer system:
| V1 |
|
| AM1 (MLT ch.1) |
| AM2 (MLT ch.2) |
| AM3 (MLT ch.3) |
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| V2 |
|
| AM4 |
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| V3 |
|
| AM5 |
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| V 4 |
| V 5 |
|
| V 6 |
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
Calibration gas mixtures |
|
|
| |||||||||||
Appropriate assignment: |
|
|
|
|
|
|
| |||||||
|
| AM1/Ch.1 |
|
| AM2/Ch.2 | AM3/Ch.3 | AM4 | AM5 | ||||||
| V1 |
|
|
| V1 | V1 | V2 | V3 | ||||||
| V4 |
|
|
| V4 | V5 | V5 | V6 | ||||||
| V5 |
|
|
| V5 | V4 | V6 | V5 | ||||||
| V5 |
|
|
| V5 | V4 | V6 | V5 | ||||||
| V6 |
|
|
| V5 | V4 | V4 | V4 | ||||||
| V6 |
|
|
| V5 | V4 | V4 | V4 | ||||||
Attention: Take in mind that any MLT channel is regarded as AM in this case!
Take care that for TFID separate valves are chosen which are not related to other AM’s or MLT channels (internal TFID valves only)!
This assigning procedure is to perform for all analyzer modules which should be included in system calibration.
To display the assigned valves and purge times for each AM we can push the softkey „View“.
If we want to exclude an AM from SYSCAL we can do this by entering an invalid valve. We can watch in the display whether an AM/channel is enabled for system calibration.
Attention: Remember to assign a system valve to an
Supplement - 8 | NGA 2000 |