NGA-2000 System Calibration

A further general parameter is "Timeout for Gastest". It is located in the menu "Test procedures".

System configuration and diagnostics...

System calibration...

Calibration/Test Procedures...

Test procedures...

-- Test Procedures --

 

Gastest of specific module

SAMPLE-Gas

Timeout for Gastest (0 = no timeout)

180 s

Choose specific analyzer module...

 

_______________________________________________

Procedure time:10 s

Analyzer-module:TFID

SCTESTGAS SCTIMEOUT

SCCALTIME1 SCSTEPMOD

This parameter is related to the possibility to activate a gas valve of a specific module for test purposes. Here we can determine a time after which an activated gastest automatically switches back onto SAMPLE-Gas.

Is this parameter set to "0" no automatic back switching is done.

Supplement - 12

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98