NGA-2000 System Calibration

3.3 Set-up general parameters

One general parameter is "Calibration Procedures in Test-Mode". It is located in the menu "Calibration/Test procedures".

System configuration and diagnostics...

System calibration...

Calibration/Test procedures...

-- Calibration/Test Procedures --

Start Zeroing all ranges!

 

 

Start Zeroing and Spanning all ranges !

 

 

Start calibration program !

 

 

Cancel calibration!

 

 

Calibration Procedures in Test-Mode:

 

No

Test Procedures...

 

 

Calibration type:

ZERO-CAL

Program step:

1

Calibration time:

 

16 s

Previous calibration time:

 

57 s

 

 

 

 

 

Result...

SCTESTMOD

SCCALTYPE2 SCPROGSTEP SCCALTIME1 SCCALTIME2

With this parameter we can run all the calibration procedures either in the defined mode or in a test mode (set test mode to „Yes“). The test mode means that the valve switching and waiting for purge times is done in the same manner like in the normal calibration procedure.

The only difference is that the single calibrations of the modules and the appropriate times the modules would need for the calibration are not done.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98