3.2 Programming Calibration Sequences
Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to run the system calibration in an user definable order of up to 40 steps.
The menu to
System configuration and diagnostics...
↓
System calibration...
↓
Calibration sequence programming...
↓
Step #: | 1 |
Calibration procedure type: | |
Choose specific analyzer module... |
|
Select all analyzer modules! |
|
Program steps |
|
Program steps |
|
Program steps |
|
Program steps
_______________________________________________
SCSTEP SCCALTYPE1
SCSTEPMOD
For programming the sequence you must
1.select „Step #“
2.select „Calibration procedure type“
3.select analyzer module/ MLT channel
Repeat this order for all program steps.
The programmable „Calibration procedure types“ are:
1. | NoOp | no operation (for deleting a step in an existing program) |
2. | do a zero calibration | |
3. | do a span calibration for all available ranges | |
4. | do a zero cal then a span cal for all available ranges | |
5. | do a span calibration only for range #1 | |
6. | do a span calibration only for range #2 | |
7. | do a span calibration only for range #3 | |
8. | do a span calibration only for range #4 | |
9. | end of sequence |
90003482(2) | NGA 2000 | Supplement - 9 |