NGA-2000 System Calibration

3.2 Programming Calibration Sequences

Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to run the system calibration in an user definable order of up to 40 steps.

The menu to set-up this feature looks as follows.

System configuration and diagnostics...

System calibration...

Calibration sequence programming...

-- Calibration Sequence Programming --

Step #:

1

Calibration procedure type:

ZERO-CAL

Choose specific analyzer module...

 

Select all analyzer modules!

 

Program steps 1-10...

 

Program steps 11-20...

 

Program steps 21-30...

 

Program steps 31-40...

_______________________________________________

Analyzer-module:MLT/CH3

SCSTEP SCCALTYPE1

SCSTEPMOD

For programming the sequence you must

1.select „Step #“

2.select „Calibration procedure type“

3.select analyzer module/ MLT channel

Repeat this order for all program steps.

The programmable „Calibration procedure types“ are:

1.

NoOp

no operation (for deleting a step in an existing program)

2.

Zero-Cal

do a zero calibration

3.

Span-Cal

do a span calibration for all available ranges

4.

Zero/Span-Cal

do a zero cal then a span cal for all available ranges

5.

Span1-Cal

do a span calibration only for range #1

6.

Span2-Cal

do a span calibration only for range #2

7.

Span3-Cal

do a span calibration only for range #3

8.

Span4-Cal

do a span calibration only for range #4

9.

END-OF-PGRM

end of sequence

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

Supplement - 9