3.1
Before we can run any system calibration features we have to include the different analyzer modules (AM’s) into system calibration by setup needed parameters. Only after correct setting up, an AM is included into system calibration and only then it makes sense to go into the other menus.
System configuration and diagnostics...
↓
System calibration...
↓
↓
Analyzer Set up |
|
Choose analyzer module.... |
|
|
|
Gas type: | |
Assigned to system valve: | V5 |
Purge time: | 20 s |
Note: Ranges are always calibrated separately !
_______________________________________________
Analyzer module: | MLT/CH2 | |
Module enabled for | NO | |
|
|
|
| View... |
|
SCAMGAS SCVALVE SCPURGE
SCMODULE SCCONTROL
Setting up an AM for system calibration means assigning valves from system valve pool. The Control Module provides support of up to 32
decide which valves deliver which gases for an analyzer module. Also we have to know the purge time from a valve to the AM/ MLT channel.
For each of the following types in the parameter „Gas type“ we have to assign a valve and the appropriate purge time :
∙
∙
∙
∙
∙
∙
Conditions for the valve assignment:
·Once a valve has been assigned to be a sample gas valve for any AM it must not be used for zero or span gases !
·The zero valve of an AM can not be a span valve of the same AM.
·All gas types have to be assigned to a
·TFID has internal valves to be assigned separately as system valves!
90003482(2) | NGA 2000 | Supplement - 7 |