4.5Zero Calibration

Change to the menu "Analyzer Module Calibration"

F5

Press

Caution:

Before starting zero calibration, realize that zero gas is available ! (See also section 5.1.1, p. 5-5... !)

Note:

The zeroing of all measurement ranges of the same channel is running simulta- neously.

Change to the line "Start zero calibration procedure !"

Press once to get the line "Start zero calibration procedure !" white on black.

Start the zero calibration

Press

or

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

4 - 39