TAG | 37.50 ppm |
Do you really want to do this ??
Press "Yes" or "Back..."
YesBack...
F1 | F2 | F3 | F4 | F5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| TAG |
|
|
|
|
|
|
|
|
|
|
| 37.50 ppm |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Procedure status: |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Remaining procedure time: |
|
|
|
|
|
| 8 s |
|
|
|
|
|
|
| |||||||
| Current/expected gas flow: |
|
|
|
|
| Zerogas |
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Concentration in span gas units: |
|
|
|
| 37.50 ppm |
|
|
|
|
|
|
| |||||||||
| Last zero |
| Results |
|
|
|
| Success |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| Last span calibration: |
|
|
|
|
|
|
|
| Success |
|
|
|
|
|
|
| |||||
| Last zero calibration was: | Fri |
|
|
|
|
|
|
| |||||||||||||
| Last span calibration was: | Fri |
|
|
|
|
|
|
| |||||||||||||
| Successful zero+span calibrated ranges: | 1+2+3+4 |
|
|
|
|
|
|
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Measure |
| Cancel ! |
|
|
|
| Back... |
| More... |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 | F3 |
| F4 |
|
| F5 |
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| TAG |
|
|
|
|
|
|
|
|
|
|
| 3.13 ppm |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Procedure status: |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Remaining procedure time: |
|
|
|
|
|
| 97 s |
|
|
|
|
|
|
| |||||||
| Current/expected gas flow: |
|
|
|
|
| Zerogas |
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Concentration in span gas units: |
|
|
|
| 3.13 ppm |
|
|
|
|
|
|
| |||||||||
| Last zero calibration: |
| Results |
|
|
|
| Success |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| Last span calibration: |
|
|
|
|
|
|
|
| Success |
|
|
|
|
|
|
| |||||
| Last zero calibration was: | Fri |
|
|
|
|
|
|
| |||||||||||||
|
|
|
|
|
|
|
| |||||||||||||||
| Last span calibration was: | Fri |
|
|
|
|
|
|
| |||||||||||||
| Successful zero+span calibrated ranges: | 1+2+3+4 |
|
|
|
|
|
|
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Measure |
| Cancel ! |
|
|
|
| Back... |
| More... |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
|
| F2 |
|
| F3 |
|
| F4 |
|
|
| F5 |
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
⇒Confirm the order for zero calibration
Press | F2 |
to start |
the zeroing immediately.
Option:
Press the F4 key if you want to cancel the order.
Notes:
·The display of this message depends on the setup in the expert configurations. (see 5.1.8 p.
·The 3 following illustrations show the displays you can see after starting the zeroing procedure.
⇒Zeroing:
1.
Notes:
·The procedure time de- pends on the parameters entered for purge time (see section 5.1.1 p.
·The purge time must be long enough to get a stable signal before calibration.
·You can cancel the run- ning calibration procedure at any time with the F2 key.
⇒Zeroing:
2.
Note:
The procedure time depends on the parameters entered for stability time and averaging time.
4 - 40 | NGA 2000 |
90003482(2)