Saving the factors:
♦You can store the actual zero offset and span factor with the F2 key (STORE) of the menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu page (see next illustration).
♦You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with the F4 key (HISTORY):
CLD |
|
|
| 95.0 ppm |
|
| Range 3 Factors |
| |
Zero offset: | Manufacturer´s settings. | 0.000000 | ||
|
|
| ||
Span factor: |
|
|
| 0.000000 |
Zero offset: |
| Stored settings |
| 0.000000 |
|
|
| ||
Span factor: |
|
|
| 0.000000 |
HOME | NEXT | RSTR MN | RSTR ST | INFO |
F1 | F2 | F3 | F4 | F5 |
♦You can change the stored settings against the manufacturer's settings with the F3 key (RSTR MN: Restore the manufacturer's values). The "Zero offset" and the "Span factor" of the first menu page will change according to the setups of the manufacturer's settings.
♦You can change the manufacturer's settings against the stored settings with the F4 key (RSTR ST: Restore the "stored" values). The "Zero offset" and the "Span factor" of the first menu page will change according to the setups of the stored settings.
90003749(1)
NGA 2000 | 5 - 59 |