5.5.3 Expert Analyzer Controls
Physical Measurements
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Expert analyzer controls | ||||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Physical Measurements | ||||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| CLD |
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| ||||
|
|
|
|
| Physical Measurements |
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| Bypass sample flow: |
|
|
|
|
| 1100 ml/min |
|
|
|
|
|
|
| |||||||||
| Flow lower limit: |
|
|
|
|
| 200 | ml/min |
|
|
|
|
|
|
|
| |||||||
| Flow upper limit: |
|
|
|
|
| 1500 | ml/min |
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||||
| Sample pressure: |
|
|
|
|
|
| 340 hPa |
|
|
|
|
|
|
| ||||||||
| Ozone supply pressure: |
|
|
|
|
| 1030 hPa |
|
|
|
|
|
|
| |||||||||
| Sensor block temperature: |
|
|
|
|
| 0.5 C |
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| NO/NOx |
|
|
|
|
| INFO |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| F1 |
| F2 | F3 | F4 |
| F5 |
| |||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Physical Measurements" you can find some pressure results and temperature results of the CLD analyzer module. Besides, you can set up flow limits. You can select the gas component with the F3 key (NO/NOx).
Set up parameters:
¨Select the line "Flow lower limit" or "Flow upper limit" with the −
¨Press the ENTER key or the →
¨Select any digit with the ←
−
Options: | |
| 0 to 2000 ml/min for the upper limit |
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Lines of variables "Bypass sample flow", "Sample pressure", "Ozone supply pressure" and "Sensor block temperature":
These are information lines. They are not available.
Lines of variables "Flow lower limit", "Flow upper limit":
The violation of these limits may start an alarm report, if it is enabled in the diagnostic menus (see 6.3 p