5.5.3 Expert Analyzer Controls

Physical Measurements

Main Menu — Expert controls and set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Expert analyzer controls

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Physical Measurements

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Physical Measurements

 

 

 

 

 

 

 

 

 

 

 

 

 

Bypass sample flow:

 

 

 

 

 

1100 ml/min

 

 

 

 

 

 

 

 

Flow lower limit:

 

 

 

 

 

200

ml/min

 

 

 

 

 

 

 

 

 

Flow upper limit:

 

 

 

 

 

1500

ml/min

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sample pressure:

 

 

 

 

 

 

340 hPa

 

 

 

 

 

 

 

 

Ozone supply pressure:

 

 

 

 

 

1030 hPa

 

 

 

 

 

 

 

 

Sensor block temperature:

 

 

 

 

 

0.5 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

NO/NOx

 

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Physical Measurements" you can find some pressure results and temperature results of the CLD analyzer module. Besides, you can set up flow limits. You can select the gas component with the F3 key (NO/NOx).

Set up parameters:

¨Select the line "Flow lower limit" or "Flow upper limit" with the -key or the -key.

¨Press the ENTER key or the -key to select the parameter.

¨Select any digit with the -key or the -key and adjust a new value with the

-key or the -key

Options:

-200 to +2000 ml/min for the lower limit (sign changing with F4 "+/-")

 

0 to 2000 ml/min for the upper limit

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

Lines of variables "Bypass sample flow", "Sample pressure", "Ozone supply pressure" and "Sensor block temperature":

These are information lines. They are not available.

Lines of variables "Flow lower limit", "Flow upper limit":

The violation of these limits may start an alarm report, if it is enabled in the diagnostic menus (see 6.3 p 6-27).

5 - 60

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98