CLD |
|
|
| 95.0 ppm |
|
| Range 3 Factors |
| |
Zero offset: |
|
|
| 521580.6 |
Span factor: |
|
| 0.002608716 | |
Full scale range at calibration: |
| 100.0 ppm | ||
Measurement range number: |
| 3 | ||
Raw measurement signal: |
| 521660 | ||
HOME | STORE | NEXT | HISTORY | INFO |
F1 | F2 | F3 | F4 | F5 |
Set up parameters:
¨Press the ENTER key or the →
Note: The zero offset is about 32000 counts for the CLD software revision 2.2.1 (16 bit D/A transformer) and about 520000 counts for the revision 2.3 (20 bit D/A transformer).
¨Confirm your setup with the ENTER key.
¨Go to the line "Span factor" with the ↓
¨Press the ENTER key or the →
¨Confirm your setup with the ENTER key.
Notes:
¨You have to set up the factors for all ranges.
¨With the parameter "Zero offset" you will fix the zero value. The changing of the zero offset will influence the signal immediately (look at the headline).
¨With the parameter "Span factor" you will fix the signal at the span gas setpoint. The changing of the span factor will influence the signal immediately (look at the headline).
¨Conditions for the manual input of factors:
·"Disabled" has to be set up in the line "Calibration adjustment limits" of the menu "Zero/span calibration".
·"SEPARATELY" has to be set up in the lines "Zero ranges" and "Span ranges" of the menu "Calibration Parameters" (see 5.1.2, p.
¨Attention:
After the calibration with factors you have to calibrate with zero gas and span gas in the usual way!
5 - 58 | NGA 2000 |