5.5.2 Expert Analyzer Controls
Zero/Span Calibration
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Expert analyzer controls | ||||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Zero/Span calibration | ||||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
| Zero/span calibration |
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Measurement range number: |
|
|
| 3 |
|
|
|
|
|
|
|
| ||||||||||
| Zero gas concentration: |
|
|
|
|
|
|
| 0.0 ppm |
|
|
|
|
|
|
|
| ||||||
| Span gas concentration: |
|
|
|
|
|
| 100.0 ppm |
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Sample flow: |
|
|
|
|
| 1100 ml/min |
|
|
|
|
|
|
| |||||||||
| Raw measurement signal: |
|
|
| 521590 |
|
|
|
|
|
|
|
| ||||||||||
| Measurement gas: |
|
|
|
|
|
|
|
| NOx |
|
|
|
|
|
|
| ||||||
| NO/NOx toggle! |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Status: |
|
|
|
|
|
|
|
|
|
|
| READY |
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Result... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Calibration adjustment limits: |
|
|
|
|
| Disabled |
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| FACTORS |
| ZERO |
| SPAN |
|
| INFO |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| F1 |
|
| F2 |
|
| F3 |
|
| F4 |
|
|
| F5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Zero/span calibration" you can set up and start the calibration of the analyzer. The procedure and the menus are the same as described for the analyzer module in chapter 5.1.1 (p.
How to calibrate:
1)Select the measurement range:
·Press the ENTER key or the →
2)Select the gas component:
·Go to the line "NO/NOx toggle!" with the ↓
3)Zeroing:
·Change to the submenu "Analyzer zero" with the F3 key (ZERO).
·Pay attention to the notes on the screen !
·Flow enough zero gas through the CLD.
·Wait for a stable signal (look at the headline of the display). Then, start the zero gas calibration with the F3 key (ZERO). The running procedure will be the same as it is described detailed in chapter 4.2.1 of the basic controls (p.
·Go back to the menu "Zero/span calibration" with the F2 key or the ←
5 - 56 | NGA 2000 |
90003749(1)