CLD | 95.0 ppm |
Analyzer zero
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any
| If you are sure, press ZERO again now. |
|
| ||||||||||
| Press the left arrow key when you are done. |
|
| ||||||||||
| Calibration status: |
|
|
|
|
|
| READY | |||||
| Error message for last zero: |
|
|
| CAL OK | ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| ZERO |
|
|
| INFO | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| F1 |
| F2 |
| F3 | F4 | F5 |
CLD | 13.7 ppm |
Analyzer zero
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any
If you are sure, press ZERO again now. Press the left arrow key when you are done.
Calibration status: |
|
| ZEROING - WAIT | |
Error message for last zero: |
| CAL OK | ||
| ABORT |
|
|
|
F1 | F2 | F3 | F4 | F5 |
CLD | 0.057 ppm |
Analyzer zero
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any
| If you are sure, press ZERO again now. |
|
| ||||||||||
| Press the left arrow key when you are done. |
|
| ||||||||||
| Calibration status: |
|
|
|
|
|
| READY | |||||
| Error message for last zero: |
|
|
| CAL OK | ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| ZERO |
|
|
| INFO | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| F1 |
| F2 |
| F3 | F4 | F5 |
⇒Start Zeroing
+ Pay attention to the note at p.
F3
Press
Caution:
You have to
·flow enough zero gas through the analyzer
·look for a stabilized signal
before starting the zero gas calibration.
Note:
You can go back to the menu "Basic Controls" with the F2 key.
⇒Zero gas calibration is running
Note:
You can abort the procedure at any time with the F2 key.
⇒Change to the main menu
F1
Press
Alternative:
If you want to continue with the span gas calibration after the zeroing is done, you can go back to the basic controls with the F2 key or the ←
(The spanning will be described in the next chapter: 4.2.2).
4 - 22 | NGA 2000 |
90003749(1)