4.2.1 Calibration - Zeroing

CLD

 

 

 

0.017 ppm

 

 

Main Menu

 

 

Basic controls...

 

 

 

Expert controls and set up...

 

 

(Operational configuration)

 

 

Technical level configuration...

 

 

(Diagnostic and manufacturing/service)

 

MEASURE

STATUS

NEXT

LOCK

INFO

F1

F2

F3

F4

F5

CLD

 

 

 

 

 

 

 

 

 

 

 

0.017

 

 

 

 

 

 

 

ppm NOx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

Range: 4

250

Sample press.:

340 hPa

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector temperature: 0.5

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Block temperature:

51.5

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Converter temp.:

360

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DISPLAY STATUS MENU NEXT INFO

F1

F2

F3

F4

F5

Notes for zero gas calibration:

·You can calibrate all ranges together or each range separately with zero gas.

·You have to set up the parameter you want in the menu "Calibration Parameters" (see chapter 5.1.2 S. 5-9).

Change to the single component display

F1

Press

Single component display will appear after the zero gas calibration

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000

4 - 23