4.2.2 Calibration - Spanning

CLD

 

 

 

225.0 ppm

 

 

Main Menu

 

 

Basic controls...

 

 

 

Expert controls and set up...

 

 

(Operational configuration)

 

 

Technical level configuration...

 

 

(Diagnostic and manufacturing/service)

 

MEASURE

STATUS

NEXT

LOCK

INFO

F1

F2

F3

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

225.00

 

 

 

 

 

 

 

ppm NOx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

Range: 4

 

 

 

250

 

 

 

 

 

 

 

 

 

 

Sample press.:

 

340 hPa

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector temperature:

0.5

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Block temperature:

 

 

51.5

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Converter temp.:

 

 

360

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DISPLAY

 

STATUS

 

MENU

 

NEXT

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

 

 

F2

 

 

 

F3

 

 

 

F4

 

 

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Change to the single component display

F1

Press

Single component display will appear after the span gas calibration

Notes for the span gas calibration:

·You can calibrate all ranges together or each range separately with span gas. You have to set up the parameter you want in the menu "Calibration Parameters" (see chapter 5.1.2 p. 5-9).

·If it will not be possible to calibrate all ranges with the same span gas, you have to calibrate each range separately!

·The desired value of span gas has to be (!) a value between 10% and 110% of each end of range. Otherwise, no calibration will be running.

Example:

· End of range 1: 50 ppm

·End of range 2: 250 ppm

·End of range 3: 1000 ppm

·End of range 4: 2500 ppm

®

Potential span gas setpoints:

5

55 ppm

®

Potential span gas setpoints:

25

275 ppm

®Potential span gas setpoints: 100 – 1100 ppm

®Potential span gas setpoints: 250 – 2750 ppm

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98