| CLD |
|
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
| |||
|
|
|
|
|
| Basic Controls |
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| Measurement range number: |
|
|
| 4 |
|
|
|
|
|
|
| ||||||||||
| Range upper limit: |
|
|
|
|
|
|
| 250.0 ppm |
|
|
|
|
|
|
| ||||||
| Range and functional control: |
|
|
|
| LOCAL |
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
| Ranges with valid calibration: |
|
|
| 4 |
|
|
|
|
|
|
| ||||||||||
| Calibration status: |
|
|
|
|
|
|
|
| READY |
|
|
|
|
|
| ||||||
| If it won´t calibrate... |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Status: |
|
|
|
|
|
|
|
|
|
| STANDBY |
|
|
|
|
|
|
| |||
| Measurement mode: |
|
|
|
|
|
|
|
|
| NOx |
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Ozonator status: |
|
|
|
|
|
|
|
|
| On |
|
|
|
|
|
| |||||
| Ozonator power: |
|
|
|
|
|
|
|
| Enabled |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| NO/NOx |
| ZERO |
| SPAN |
| INFO |
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 |
| F3 | F4 |
| F5 | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
CLD | 95.0 ppm |
If it won´t calibrate...
Check that you are flowing the correct gas, and the gas concentration is what it is supposed to be. Make sure that the reading is stable before starting. If you have changed the range full scale value, or any linearizer coefficients, or enabled or disabled it,
or done anything else that would affect how it measures the gas, you may have made it hard for the algorithm to get to a calibration.
In this case, manually adjust the coefficients
until the readings are close to correct, and try again.
HOME | ESCAPE |
|
| INFO |
F1 | F2 | F3 | F4 | F5 |
CLD |
|
|
| 95.0 ppm |
|
| Main Menu |
|
|
Basic controls... |
|
|
| |
Expert controls and set up... |
|
| ||
(Operational configuration) |
|
| ||
Technical level configuration... |
|
| ||
(Diagnostic and manufacturing/service) |
| |||
MEASURE | STATUS | NEXT | LOCK | INFO |
F1 | F2 | F3 | F4 | F5 |
⇒Change to the menu page "If it won´t calibrate..."
Press
Notes:
·Follow the instructions on the screen
to calibrate succesfully!
·If you have further questions, contact your customer service, please!
⇒Change to the main menu
F1
Press
Alternative:
If you want to continue with further setups of the basic controls, you can go back to the corresponding menu with the F2 key or the ←
⇒Change to the single component display
F1
Press
4 - 30 | NGA 2000 |
90003749(1)