5.1.1 Analyzer Module Set up
Calibration/Calibration Gas List
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
| Analyzer module set up | ||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
| Calibration gas list | ||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
| Calibration Gas List |
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| Zero gas - range 1: |
|
|
|
|
| 0.00 ppm |
|
|
|
|
|
|
|
| ||||||
| NO Span gas - range 1: |
|
|
|
|
| 2.00 ppm |
|
|
|
|
|
|
|
| ||||||
| NOx Span gas - range 1: |
|
|
|
|
| 2.00 ppm |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Zero gas - range 2: |
|
|
|
|
| 0.00 ppm |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| NO Span gas - range 2: |
|
|
|
| 25.00 ppm |
|
|
|
|
|
|
| ||||||||
| NOx Span gas - range 2: |
|
|
|
| 25.00 ppm |
|
|
|
|
|
|
| ||||||||
| Calibration... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| MORE |
|
|
| INFO |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| F1 |
| F2 | F3 | F4 |
| F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the second page you will find the parameters of range 3 and 4. You can reach the second menu page with the F3 key (MORE).
Via the menu line "Calibration..." you can change to the menu "Zero/span calibration", where you can start zeroing and spanning (see next page).
Set up parameters:
¨Select the line of variables you want or the menu line "Calibration..." with the −
¨Select the parameter or change to the menu "Zero/span calibration" with the ENTER key or the →
¨Select any digit with the ←
the −
Options zero gas: | 0.00 ppm up to the end of range value |
Options span gas: | 0.00 ppm up to 120% of the end of range value |
| You can set up the end of range values |
| in the menu "Range Settings" (see 5.1.6 p. |
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
5 - 4 | NGA 2000 |
90003749(1)