5.1.1 Analyzer Module Set up

Calibration/Calibration Gas List

Main Menu — Expert controls and set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration gas list

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Calibration Gas List

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero gas - range 1:

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

 

NO Span gas - range 1:

 

 

 

 

 

2.00 ppm

 

 

 

 

 

 

 

 

 

NOx Span gas - range 1:

 

 

 

 

 

2.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero gas - range 2:

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NO Span gas - range 2:

 

 

 

 

25.00 ppm

 

 

 

 

 

 

 

 

NOx Span gas - range 2:

 

 

 

 

25.00 ppm

 

 

 

 

 

 

 

 

Calibration...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

MORE

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the second page you will find the parameters of range 3 and 4. You can reach the second menu page with the F3 key (MORE).

Via the menu line "Calibration..." you can change to the menu "Zero/span calibration", where you can start zeroing and spanning (see next page).

Set up parameters:

¨Select the line of variables you want or the menu line "Calibration..." with the -key or the -key.

¨Select the parameter or change to the menu "Zero/span calibration" with the ENTER key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with

the -key or the -key.

Options zero gas:

0.00 ppm up to the end of range value

Options span gas:

0.00 ppm up to 120% of the end of range value

 

You can set up the end of range values

 

in the menu "Range Settings" (see 5.1.6 p. 5-16).

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.