Span gas calibration notes:
¨The span gas setpoint of NO and NOx have to be the same value, because both components will be calibrated simultaneously!
¨You can calibrate all ranges together or each range separately with span gas. You have to set up the parameter you want in the menu "Calibration Parameters" (see chapter 5.1.2 p.
¨If it will not be possible to calibrate all ranges with the same span gas, you have to calibrate each range separately!
¨The desired value of span gas has to be (!) a value between 10% and 110% of each end of range. Otherwise, no calibration will be running:
· Example: |
|
|
|
|
|
|
End of range 1: | 50 ppm | ® Potential span gas setpoints: | 5 | – | 55 ppm | |
End of range 2: | 250 ppm | ® Potential span gas setpoints: | 25 | – | 275 ppm | |
End of range 3: | 1000 ppm | ® | Potential span gas setpoints: | 100 | – 1100 ppm | |
End of range 4: | 2500 ppm | ® | Potential span gas setpoints: | 250 | – 2750 ppm |
How to calibrate:
Change to the menu "Zero/span calibration" via the menu line "Calibration...":
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
|
| |||
|
|
|
|
| Zero/span calibration |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Measurement range number: |
|
|
| 3 |
|
|
|
|
|
|
|
|
| ||||||||||
| Zero gas concentration: |
|
|
|
|
|
|
| 0.0 ppm |
|
|
|
|
|
|
|
|
| ||||||
| Span gas concentration: |
|
|
|
|
|
| 100.0 ppm |
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Sample flow: |
|
|
|
|
| 1100 ml/min |
|
|
|
|
|
|
|
| |||||||||
| Raw measurement signal: |
|
|
| 521590 |
|
|
|
|
|
|
|
|
| ||||||||||
| Measurement gas: |
|
|
|
|
|
|
|
| NOx |
|
|
|
|
|
|
|
| ||||||
| NO/NOx toggle! |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Status: |
|
|
|
|
|
|
|
|
|
|
| READY |
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Result... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Calibration adjustment limits: |
|
|
|
|
| Disabled |
|
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| FACTORS |
| ZERO |
| SPAN |
|
| INFO |
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| F1 |
|
| F2 |
|
| F3 |
|
| F4 |
|
|
| F5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1)Select the measurement range:
·Press the ENTER key or the →
2)Select the gas component:
·Go to the line "NO/NOx toggle!" with the ↓
90003749(1)
NGA 2000 | 5 - 5 |