5.1.2 Analyzer Module Set up
Calibration Parameters
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| Analyzer module set up | |||||||||||||||||
|
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| Calibration parameters | |||||||||||||||||
|
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
| Calibration Parameters |
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Calibration adjustment limits: |
|
|
|
| Disabled |
|
|
|
|
|
|
|
| |||||||||
| Calibration averaging time: |
|
|
|
|
| 5 s |
|
|
|
|
|
|
|
| ||||||||
| Calibration failure alarm: |
|
|
|
|
| No |
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
| Cal. failure error allowed: |
|
|
| 50 % |
|
|
|
|
|
|
|
| ||||||||||
| Calibration time out: |
|
|
|
|
|
|
|
|
| 60 s |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Zero ranges: |
|
|
|
|
|
|
| TOGETHER |
|
|
|
|
|
|
| |||||||
| Span ranges: |
|
|
|
|
|
|
| SEPARATELY |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
|
|
|
|
|
|
|
| INFO |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 |
| F3 | F4 |
| F5 |
| ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Calibration Parameters" you can set up several parameters of the zero and span calibration.
Set up parameters:
¨Select the line of variables you want with the ↓
¨Select the variable with the ENTER key or the →
¨Select any digit with the ←
resp. select the whole parameter with the −
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Line of variables "Calibration adjustment limits:
¨Enabled: The calibration result will only be accepted within the limit settings. If the results will be out of these limits, you will find a corresponding error message in the menu "Analyzer zero" respectively "Analyzer span". Illustrations of these menus you will find in chapter 4.2.1 p.
¨Disabled: The calibration will be possible, even if the calibration results will leave the allowed limits. Compare with the explanations in chapter 5.1.1 p.