5.1.4 Analyzer Module Set up
Linearization Parameters
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| Analyzer module set up | |||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Gas measurement parameters | ||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| Linearization parameters | |||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
| Linearization parameters |
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Range 1 linearizer: |
|
|
|
|
|
| Disabled |
|
|
|
|
|
|
|
| |||||
| If enabled, uses curve no.: |
|
|
| 1 |
|
|
|
|
|
|
|
| ||||||||
| Range 2 linearizer: |
|
|
|
|
|
| Disabled |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| If enabled, uses curve no.: |
|
|
| 2 |
|
|
|
|
|
|
|
| ||||||||
| Range 3 linearizer: |
|
|
|
|
|
| Disabled |
|
|
|
|
|
|
| ||||||
| If enabled, uses curve no.: |
|
|
| 3 |
|
|
|
|
|
|
|
| ||||||||
| Range 4 linearizer: |
|
|
|
|
|
| Disabled |
|
|
|
|
|
|
| ||||||
| If enabled, uses curve no.: |
|
|
| 4 |
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
| |||||||||||
| Set coefficients... |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| HOME |
| ESCAPE |
|
|
|
|
| INFO |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 |
| F3 | F4 | F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Linearization parameters" you can enable the linearization for each range and you can select the linearization curve.
Note:
Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.
Set up parameters:
¨Select any line of variables or the menu line "Set coefficients..." with the ↓
¨Select the variable or change to the submenu "Linearity coefficients" with the ENTER key or the →
¨Select the parameter you want with the −
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.