5.1.4 Analyzer Module Set up

Linearization Parameters

Main Menu — Expert controls and set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gas measurement parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Linearization parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Linearization parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Range 1 linearizer:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

 

If enabled, uses curve no.:

 

 

 

1

 

 

 

 

 

 

 

 

 

Range 2 linearizer:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If enabled, uses curve no.:

 

 

 

2

 

 

 

 

 

 

 

 

 

Range 3 linearizer:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

If enabled, uses curve no.:

 

 

 

3

 

 

 

 

 

 

 

 

 

Range 4 linearizer:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

If enabled, uses curve no.:

 

 

 

4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set coefficients...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Linearization parameters" you can enable the linearization for each range and you can select the linearization curve.

Note:

Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.

Set up parameters:

¨Select any line of variables or the menu line "Set coefficients..." with the -key or the -key.

¨Select the variable or change to the submenu "Linearity coefficients" with the ENTER key or the -key.

¨Select the parameter you want with the -key or the -key.

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

5 - 12