5.1.8 Analyzer Module Set up

Linearization Functions

Main Menu — Expert controls and set up

Analyzer module set up

Gas measurement parameters

Linearization functions

CLD

95.0 ppm

Linearization functions

Polynomial set up...

Midpoint correction set up...

Use the polynomial set up to generate a linearizing polynomial from up to 20 gases. With more than 6 gases it will produce a fourth order polynomial linearizer.

Use the midpoint correction for a piecewise-linear final correction, to bring up to three points precisely onto the curve.

HOME

ESCAPE

 

 

INFO

F1

F2

F3

F4

F5

From the menu "Linearization functions" you can change to several submenus, where you can linearize the ranges and optimize the linearization functions:

¨Change to the line you want with the -key or the -key. Note:

Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.

Polynomial set up:

Press the ENTER key or the -key in the line "Polynomial set up..." to change to the corresponding menu:

5 - 18

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98