5.1.8 Analyzer Module Set up
Linearization Functions
Main Menu — Expert controls and set up
↓
Analyzer module set up
↓
Gas measurement parameters
↓
Linearization functions
↓
CLD | 95.0 ppm |
Linearization functions
Polynomial set up...
Midpoint correction set up...
Use the polynomial set up to generate a linearizing polynomial from up to 20 gases. With more than 6 gases it will produce a fourth order polynomial linearizer.
Use the midpoint correction for a
HOME | ESCAPE |
|
| INFO |
F1 | F2 | F3 | F4 | F5 |
From the menu "Linearization functions" you can change to several submenus, where you can linearize the ranges and optimize the linearization functions:
¨Change to the line you want with the ↓
Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.
Polynomial set up:
Press the ENTER key or the →
5 - 18 | NGA 2000 |
90003749(1)