Line of variables "Ozonator power":

Options: Enabled/Disabled

The setup of this line will change the displayed message of the line "Ozonator status". That line is not available. The ozonator power setup is also possible in the menu "Basic Controls" (see chapter 4.1.5 p. 4-15).

Lines of variables "First/Second/Third/Fourth line's parameter:

In these lines you can set up which of the following parameters shall be displayed in each of the last four lines of the single component display:

Sample pressure (bargraph with measured value)

Sample flow (bargraph with measured value)

Raw signal (bargraph with measured value)

Validity (report: valid/invalid)

Interference (report: On/Off)

Health (report: Normal/...)

Operational state (report: STANDBY/...)

Measurement mode (report: 1/...)

Linearizer (report: Enabled/Disabled)

Calibration status (report: READY/...)

Output delay time (bargraph with value selected)

t90 time (bargraph with value selected)

Noise level (of concentration)

NO/NOx (Display: NO/NOx)

Detector temperature (bargraph with measured value)

Block temperature (bargraph with measured value)

Converter temperature (bargraph with measured value)

Ozonator (report: OFF - ENABLED/OFF - DISABLED)

Menu line "Linearizer parameters...":

Via this line you can change to the submenu "Linearity coefficients", where you can set up the coefficients of the linearization curves manual. You can also reach the menu "Linearity coefficients" via the menu "Linearization parameters". These menus are described detailed in chapter 5.1.4 p. 5-12/13.

Ramifications via function keys:

With the F3 key (NEXT) of the menu "Analyzer Parameter List" you can change to further menu pages, where you can set up the following concentration and calibration parameters:

Control mode

Output delay time

Begin and end of all four ranges

t90 time (response time) for all ranges

Linearizer enabling/disabling for each range

Calibration averaging time