6)Correction of the linearization polynom, if necessary:

·You can optimize the calculated linearization with maximum three additional points.

·Press the ENTER key or the -key in the line "Midpoint correction set up..." of the menu "Linearization functions" to change to the corresponding submenu:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Midpoint correction set up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Range 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Correction:

 

 

 

 

 

 

 

 

DISABLED

 

 

 

 

 

 

 

 

 

Point being measured:

 

 

 

 

 

 

Point 3

 

 

 

 

 

 

 

 

 

Point 1 gas concentration:

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Point 2 gas concentration:

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

Point 3 gas concentration:

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

Point 1 reading:

 

 

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

Point 2 reading:

 

 

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

Point 3 reading:

 

 

 

 

 

-1.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span gas value:

 

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

Analyzer functioning:

 

 

 

 

 

 

READY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

SET

 

RANGE 4

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

a) Range selecting:

·Range 1 will be selected automatically after entering the menu. If you want to select another range to correct, press the F4 key (RANGE 2/3/4). The range selected will be displayed in the headline of the menu.

b) Disable the correction:

·You have to disable the correction mode before calculating the correction data. Otherwise, you will have interferences during the calculation with existing values.

·Press the ENTER key or the -key in the line "Correction" to select the parameter. Set up "DISABLED" with the -key or the -key and confirm it with the ENTER key.

c) Measured point selecting:

·Go to the line "Point being measured" with the -key.

·Press the ENTER key or the -key to select the parameter. Select "Point 1" with the the -key or the -key and confirm it with the ENTER key.

d) Set up the gas concentration for point 1:

·Go to the line "Point 1 gas concentration" with the -key.

·Press the ENTER key or the -key to select the parameter.

·Select any digit with the -key or -key and set up the value you want with the the -key or the -key. If necessary, change the sign with the F4 key ("+/-").

·Confirm your set up with the ENTER key.

e) Measuring of the gas concentrations:

·Flow the analyzer with the gas concentration set up of the line "Point 1 gas concentration" and measure the raw value.

·After the signal will be stable press the F3 key (SET): The current signal will be displayed in the line "Point 1 reading". But, the analyzer will equate the raw value with the desired value of this point automatically.

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98