6)Correction of the linearization polynom, if necessary:
·You can optimize the calculated linearization with maximum three additional points.
·Press the ENTER key or the →
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| ||||
|
|
|
|
| Midpoint correction set up |
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
| Range 3 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Correction: |
|
|
|
|
|
|
|
| DISABLED |
|
|
|
|
|
|
|
| ||||
| Point being measured: |
|
|
|
|
|
| Point 3 |
|
|
|
|
|
|
|
| ||||||
| Point 1 gas concentration: |
|
|
|
|
|
|
|
|
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
| Point 2 gas concentration: |
|
|
|
|
|
|
|
|
|
| |||||||||||
| Point 3 gas concentration: |
|
|
|
|
|
|
|
|
|
| |||||||||||
| Point 1 reading: |
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Point 2 reading: |
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Point 3 reading: |
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Span gas value: |
|
|
|
|
|
| 0.00 ppm |
|
|
|
|
|
|
| |||||||
| Analyzer functioning: |
|
|
|
|
|
| READY |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| SET |
| RANGE 4 |
| INFO |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| F1 | F2 | F3 |
| F4 |
| F5 |
| ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
a) Range selecting:
·Range 1 will be selected automatically after entering the menu. If you want to select another range to correct, press the F4 key (RANGE 2/3/4). The range selected will be displayed in the headline of the menu.
b) Disable the correction:
·You have to disable the correction mode before calculating the correction data. Otherwise, you will have interferences during the calculation with existing values.
·Press the ENTER key or the →
c) Measured point selecting:
·Go to the line "Point being measured" with the ↓
·Press the ENTER key or the →
d) Set up the gas concentration for point 1:
·Go to the line "Point 1 gas concentration" with the ↓
·Press the ENTER key or the →
·Select any digit with the ←
·Confirm your set up with the ENTER key.
e) Measuring of the gas concentrations:
·Flow the analyzer with the gas concentration set up of the line "Point 1 gas concentration" and measure the raw value.
·After the signal will be stable press the F3 key (SET): The current signal will be displayed in the line "Point 1 reading". But, the analyzer will equate the raw value with the desired value of this point automatically.
90003749(1)
NGA 2000 | 5 - 21 |