CLD |
|
|
| 95.0 ppm |
| Polynomial set up |
| ||
Range to be linearized: |
|
| 3 | |
Current span gas: |
|
| 100.0 ppm | |
Calculated polynomial order: |
| 4 | ||
Gas values shown as: |
|
| ppm | |
Gas concentrations... |
|
|
| |
Analyzer function: |
|
| READY | |
HOME | ESCAPE | CALC |
| INFO |
F1 | F2 | F3 | F4 | F5 |
1)Select the range to be linearized:
·Press the ENTER key or the →
·Select the range you want with the −
2)Zeroing and Spanning of the range selected:
·e.g.: Change to the "Main Menu" with the F1 key. Change after that via the line "Basic controls..." to the corresponding menu. Press the F3 key(ZERO) resp. the F4 key (SPAN) to start the calibration (Look at the descriptions in the chapters 4.2.1 p.
·Alternative: Change to the menu "Zero/span calibration" of the "Expert controls" to start the calibration (see 5.1.1 p.
3)Determine the display of gas concentration:
·Press the ENTER key or the →
·Select the concentration setup you want with the −
·"ppm" for absolute concentrations
·"Percent of span gas" for relative concentrations, i.e. the gas value will be displayed as percentage of the span gas. This will be necessary, if you will use diluted span gas.
4)How to get the
·Press the ENTER key or the →
90003749(1)
NGA 2000 | 5 - 19 |