5.1.5 Analyzer Module Set up
Response Time/Delay Parameters
Main Menu — Expert controls and set up
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| Analyzer module set up | ||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Gas measurement parameters | |||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
| Response time/delay parameters | ||||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
| Response time/delay parameters |
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
| Range 1 t90 time: |
|
|
|
|
|
|
|
| 3.0 s |
|
|
|
|
|
|
|
| ||||
| Range 2 t90 time: |
|
|
|
|
|
|
|
| 3.0 s |
|
|
|
|
|
|
|
| ||||
| Range 3 t90 time: |
|
|
|
|
|
|
|
| 3.0 s |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| Range 4 t90 time: |
|
|
|
|
|
|
|
| 3.0 s |
|
|
|
|
|
|
| |||||
| LON update rate: |
|
|
|
|
|
| 10 per sec |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
| Output delay time: |
|
|
|
|
|
|
|
| 0.0 s |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
|
|
|
|
|
|
| INFO |
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
| F1 |
| F2 |
| F3 | F4 |
| F5 |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Response time/delay parameters" you can set up the response time (t90 time) of the analyzer for each range and the output delay of the signal.
Set up parameters:
¨Select any line of variables with the ↓
¨Select the variable with the ENTER key or the →
¨Select any digit with the ←
resp. select the whole parameter with the −
¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.
Lines of variables "Range 1/2/3/4 t90 time":
The t90 time is defined as the time that has to run until the analyzer displays 90% of the current concentration after a jump in the concentration level.
Options: 0 to 30 seconds for every range of a channel.
5 - 14 | NGA 2000 |
90003749(1)