Lines of variables "Range 1/2/3/4 linearizer":

In these lines you can enable the linearization for the range selected.

You have to set up "Disabled" in these lines as long as you will determine the data for linearization. This will be necessary to avoid interferences with existing values during the calculation of the linearization curve.

Lines of variables "If enabled, uses curve no.":

You can select for each range one of the four linearization curves from the menu "Linearity coefficients" (see next illustration).

Menu "Linearity coefficients":

You can change to the menu "Linearity coefficients" via the line "Set coefficients...":

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Linearity coefficients

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Curve 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A0 coefficient:

 

 

 

 

 

 

0.000000

 

 

 

 

 

 

 

 

 

A1 coefficient:

 

 

 

 

 

 

1.000000

 

 

 

 

 

 

 

 

 

A2 coefficient:

 

 

 

 

 

 

0.000000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A3 coefficient:

 

 

 

 

 

 

0.000000

 

 

 

 

 

 

 

 

 

A4 coefficient:

 

 

 

 

 

 

0.000000

 

 

 

 

 

 

 

 

 

Curve upper limit:

 

 

 

 

 

 

100.0 ppm

 

 

 

 

 

 

 

 

Curve over-range:

 

 

 

 

 

 

5.0 %

 

 

 

 

 

 

 

 

 

Curve under-range:

 

 

 

 

 

 

5.0 %

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Status:

 

 

 

 

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

NEXT

 

LAST

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Linearity coefficients" you can put in the desired coefficients for the linearization curve selected. With the F3 key (NEXT) you can scroll between the menu pages of all four linearization curves. You will find the number of the curve selected in the head line of each "Linearity coefficients" menu.

Line of variables "Curve upper limit":

In this line you should set up the end of range value for the range you want to determine the linearization curve.

Lines of variables "Curve over-range/Curve under-range":

In these two lines you can determine how many per cent the linearization curve for the range selected shall be over resp. under the range limits.

Line of variables "Status":

In this line you can enable or disable the linearization curve selected. This is the same setup as in the lines "Range 1/2/3/4 linearizer" of the menu "Linearization parameters".

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000