Structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the MLT software starting from the line "Expert controls and set up" in the "Main Menu". The way will be described with the software catchwords that you have to enter one after another to reach the corresponding submenu. At the end of the catchword listing you will find the illustration of the LCD screen. After that you can read the set up instructions and explanations. Sometimes the illustration will be completed by sketches or further menu pictures.
Example: | You want to set up the zero/span gas concentrations for all ranges of the | ||||
| analyzer module. |
|
|
| |
| Main Menu — Expert controls and set up | ||||
|
|
|
| ↓ |
|
|
| Analyzer module set up | |||
|
|
|
| ↓ |
|
|
| Calibration gas list | |||
|
|
|
| ↓ |
|
| CLD |
|
|
| 95.0 ppm |
|
| Calibration Gas List |
| ||
| Zero gas - range 1: |
|
| 0.00 ppm | |
| NO Span gas - range 1: |
|
| 2.00 ppm | |
| NOx Span gas - range 1: |
|
| 2.00 ppm | |
| Zero gas - range 2: |
|
| 0.00 ppm | |
| NO Span gas - range 2: |
|
| 25.00 ppm | |
| NOx Span gas - range 2: |
|
| 25.00 ppm | |
| Calibration... |
|
|
| |
| HOME | ESCAPE | MORE |
| INFO |
| F1 | F2 | F3 | F4 | F5 |
In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the second page you will find the parameters of range 3 and 4. You can reach the second menu page with the F3 key (MORE)...
Further explanations and instructions to set up the parameters will follow!
5 - 2 | NGA 2000 |