Line of variables "Invert output":

Select "Enabled" in the line "Invert output" if you want to invert the measurement signal at the actual digital output, otherwise select "Disabled". The output inversion can be necessary for instance, if you want to combine a failure with an alarm control.

The last five lines of variables in the menu "DIO-module outputs" only display some adjustments of the DIO output selected! They are not available:

¨"Module status:" displays the status of the current DIO board.

¨"Slot ID:" displays the tag number of the DIO board selected.

¨"Signal name:" displays the signal selected for the current DIO output.

¨"Signal level:" displays the status of the signal selected.

¨"Signal comes from:" displays the source module selected for the current DIO output.

Configuration of the DIO module inputs:

Press the ENTER key or the -key in the line "Inputs..." to change to the corresponding submenu:

CLD

 

 

 

95.0 ppm

 

- DIO-MODULE INPUTS -

 

Inputnumber:

 

 

1

Choose module...

 

 

 

Choose function...

 

 

 

Slot ID:

 

 

 

1

Signal name:

 

 

????

Signal level:

 

 

 

Off

Signal comes from:

 

 

????

EXPERT

 

>>>

 

INFO

F1

F2

F3

F4

F5

In the menu "DIO-MODULE INPUTS" you have to determine the configuration of all of the eight DIO module inputs!

DIO input configuration setup:

1) Select the input number:

¨Press the ENTER key or the -key in the line "Inputnumber" and adjust an input number (1, ..., 8) with the -key or the -key.

¨Confirm the number selected with the ENTER key.

2) Choose the module:

¨Change with the -key to the line "Choose module...".

¨Change to the submenu "ANALYZER MODULES" with the ENTER key or the -key. All modules joint to the platform or to the analyzer are available.

5 - 40

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98