¨Go to the line you want with the -key or the -key. If there are more than eight source modules available you have to press the F3 key to go to another menu page.

¨Select the tag of the module with the ENTER key or the -key:

The display will go back to the menu "DIO-MODULE OUTPUTS" automatically. The tag of the module selected will appear in the line "Signal comes from".

3) Select any signal:

¨Change to the line "Choose signal..." with the -key.

¨Press the ENTER key or the -key to change to the submenu "Signals":

CLD

95.0 ppm

Signals

NORMAL

MAINTENANCE REQUEST

FAILURE

CAL. IN PROGRESS

ZERO IN PROGRESS

SPAN IN PROGRESS

ZERO FAILED

SPAN FAILED

HOME ESCAPE MORE BACK INFO

F1

F2

F3

F4

F5

¨Go to the line you want with the -key or the -key. With the F3 key you can go to other menu pages containing further available signals:

·RANGE OVERFLOW

·RANGE UNDERFLOW

·FLOW TO LOW

·FLOW TO HIGH

·RANGE 1, 2, 3, 4

·OZONATOR OFF

·NOx MODE

·NO MODE

·and further signals, if need be:

All signals of the control module and of the analyzer modules joint to the platform or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "SIGNALS". The signals shown above will appear by using a CLD module as source module.

¨Select the signal with the ENTER key or the -key:

The display will go back to the menu "DIO-MODULE OUTPUTS" automatically. The signal selected will appear in the line "Signal name".